Specifications
- Lower voltage alarm display
- Easy to use with one-key operation
- High accuracy and repetition
- Automatic Power-off without signal for 8 seconds
- SpO2 and pulse monitoring, and histogram of pulse display
- It is light with gross weight of 50g, batteries included, small
in size, and
easy to carry
- Two AAA batteries are required for operation with low power
consumption,
and the continuous working time is 100 hours
The Scope of Applications
The Fingertip Oximeter can monitor pulse oxygen saturation and pulse
rate
through patient’s finger. This device is applicable to hospital,
home, oxygen
bar, the community medical center, alpine area and it also can be
used before
or after sports, and so on. This device is not appropriate to be
used for
continuous monitoring.
1. Display mode: LCD display
Pulse wave’s intensity display: 16-segment digital display
Low battery voltage indication: Segment LCD signal
2. Measuring range:
Pulse: 30bpm-240bpm
SpO2: 50%-99%
3. Accuracy:
SpO2: ± 2% (during 80%-99%)
±3% (during 70%-79%)
Pulse rate: ± 2bpm
4. Resolution:
SpO2: 1%
Pulse rate: 1bpm
5. Power supply requirement: 1.5V (AAA size) alkaline
battery × 2
Battery voltage: 2.6V-3.3V
6. Measurement at low perfusion:
The values of SpO2 and pulse rate can be displayed properly when
pulse saturation is at 0.5%.
7. Resistance to surrounding light:
The difference between the values measured in the condition of indoor
natural light and that of darkroom is less than ± 1%.
8. Automatic shutdown function:
The device will power off automatically in 8 seconds when there
is no
finger in the device.
9. Photosensitive components:
Red light (wavelength: 660nm) and infrared (wavelength: 940nm)
10. Dimensions: 66mm (L) × 36mm (W) ×
33mm (H)
11. Gross weight: 50g
Accessories:
1. A hanging cord
2. Two batteries
3. A pouch
4. A User Manual
|